What is stray light interference and why does it matter?

What is stray light interference and why does it matter?

Stray light interference
In a Pre-dispersive spectrometer the ambient stray light signal can represent a large fraction of the total light signal measured by the detector, thus, it can be a major source of error. While this error can be minimized by completely shielding the sample from all ambient sources of light, this often precludes the use of most reflectance and transmittance fiber optic probes and non-contact measurement heads.

In a Post-Dispersive spectrometer the ambient stray light scattered from the sample is also collected, but with the post-dispersive instrument only ambient stray light of the same wavelength as that being measured by the detector is added to the signal. Thus, the stray light signal represents a much smaller fraction of the total light signal measured by the detector resulting in an increase in instrument baseline stability.

Post-Dispersive spectrometers provide the greatest amount of sampling, or sample interface flexibility. Post-Dispersive systems allow for non-contact and remote measurements where Pre-Dispersive systems generally have problems with ambient stray light.