Home > Resource Center > Application Notes > Issues in Absolute Spectral Radiometric Calibration: Intercomparison of Eight Sources
 

ASD

  • Our blog
  • Contact Us
  • facebook
  • linked in
  • youtube
  • Products
    • FieldSpec

      • FieldSpec 4

        • FieldSpec 4 Standard-Res Spectroradiometer
        • FieldSpec 4 Hi-Res Spectroradiometer
        • FieldSpec 4 Wide-Res Spectroradiometer
      • FieldSpec 3

        • FieldSpec 3 Portable Spectroradiometer
        • FieldSpec 3 Hi-Res Portable Spectroradiometer
        • FieldSpec 3 Max Portable Spectroradiometer
        • HandHeld 2 Portable Spectroradiometer
    • TerraSpec

        • TerraSpec 4 Standard-Res Mineral Analyzer
        • TerraSpec 4 Hi-Res Mineral Spectrometer
    • QualitySpec

        • QualitySpec Pro Vis/NIR Spectrometer
        • QualitySpec 7000
    • LabSpec

      • LabSpec

        • LabSpec 2500/2600 Portable Vis/NIR Spectrometer
        • LabSpec 5000/5100 Portable Lab Instrument
      • Software

        • ASD Software

          • Indico Pro Spectral Acquisition Software
          • RS3 Spectral Acquisition Software
        • Third-party Software

          • GRAMS Suite
          • KaleidaGraph
          • The Unscrambler
        • Accessories

          Products
        • Applications
          • Remote Sensing

              • Field Spectroscopy
              • Ground Truthing
              • Spectral Remote Sensing
              • Crops and Soils Research
              • Ice Research and Snow Research
              • Landscape Ecology and Ecology Research
              • See more Remote Sensing
          • Mining

              • Mining Exploration
              • Mining Production
              • Extractive Metallurgy
              • See more Mining
          • Additional

              • Agriculture and Soil Analysis
              • Raw Material Inspection and Analysis
              • Forest Products
              • Pharmaceutical
              • Optics and Photonics
              • See more Additional
          Applications
        • Service and Support
          • SummitCAL
          • Instrument Service
          • Training
          • Technical Support
          • Goetz Instrument Program
           
          Service and Support
        • Events
          • ASD Events
          • Training
          Events
        • Resource Center
          • FAQs
          • Application Notes
           
          Resource Center
        • About Us
          • News
          • Careers
          • About Our Experts
          • Milestones
          • History of ASD
          • Management Team
          • Corporate Citizenship
          • Partners and Memberships
          • Why NIR
           
          About Us
        Resource Center
        Resource Center
          • FAQs
          • Application Notes

        Related Applications

        • Spectroradiometry and Radiometric Calibration

        Application Notes

        Issues in Absolute Spectral Radiometric Calibration: Intercomparison of Eight Sources
        Presented January 13, 1998. Seventh AVIRIS Conference, Pasadena, California

        Alexander F. H. Goetz1,2, Bruce Kindel1,
        1Center for the Study of Earth from Space/CIRES
        2Department of Geological Sciences, University of Colorado, Boulder, CO 80309
        Peter Pilewskie
        NASA Ames Research Center, Moffett Field, CA

        The application of atmospheric models to AVIRIS and other spectral imaging data to derive surface reflectance requires that the sensor output be calibrated to absolute radiance (Green et al, 1996; Chrien et al, 1996; Gao et al, 1993). Uncertainties in absolute calibration are to be expected, and claims of ±2% accuracy have been published (Chrien et al, 1996). Measurements of accurate surface albedos and cloud absorption to be used in radiative balance calculations depend critically on knowing the absolute spectral-radiometric response of the sensor. The Earth Observing System project is implementing a rigorous program of absolute radiometric calibration for all optical sensors (Butler and Johnson, 1996a) Since a number of imaging instruments that provide output in terms of absolute radiance are calibrated at different sites, it is important to determine the errors that can be expected among calibration sites. Another question exists about the errors in the absolute knowledge of the exoatmospheric spectral solar irradiance (Nekel and Labs, 1984).

        Full article link

        full article
      • Mining
      • Additional
      • Remote Sensing
      • NIR Community Blog

        • From the Alaskan Tundra to the Tip of Africa – FieldSpec Users Share Their Stories

          February 06, 2012

        • ASD Announces 2012 Alexander Goetz Instrument Support Program Winners

          January 30, 2012

        • New Look and Structure for Spectroscopy Solutions Company ASD Inc.’s Website!

          January 23, 2012

        Read All

        • Contact Us
        • Find a Representative
        • Join our NIR Community

        Products

        • FieldSpec
        • TerraSpec
        • QualitySpec
        • LabSpec
        • Software
        • Spectroscopy Accessories

        Applications

        • Remote Sensing
        • Mining
        • Additional

        Service and Support

        • SummitCAL
        • Instrument Service
        • Training
        • Technical Support
        • Goetz Instrument Program

        Events

        Resource Center

        • FAQs
        • Application Notes

        About Us

        • News
        • Careers
        • About Our Experts
        • Milestones
        • History of ASD
        • Management Team
        • Corporate Citizenship
        • Partners and Memberships
        • Why NIR

        Copyright 2011 by ASD Inc. Terms & Conditions | Privacy Policy | Site Map | Support

        Connect with ASD Inc.: www.facebook.com www.youtube.com